Operational Validation of the Air Quality Monitor on the International Space Station

dc.creatorLimero, Thomas
dc.creatorWallace, William
dc.creatorJames, John T.
dc.date.accessioned2014-10-22T15:36:16Z
dc.date.available2014-10-22T15:36:16Z
dc.date.issued2014-07-13
dc.descriptionTucson, Arizona
dc.descriptionThomas Limero, Wyle Science, Technology, and Engineering Group, USA
dc.descriptionWilliam Wallace, Wyle Science, Technology, and Engineering Group, USA
dc.descriptionJohn T. James, Wyle Science, Technology, and Engineering Group, USA
dc.descriptionThe 44th International Conference on Environmental Systems was held in Tuscon, Arizona, USA on 13 July 2014 through 17 July 2014.
dc.description.abstractTwo air quality monitors (AQMs) were launched to the International Space Station (ISS) in March 2013. The AQM units, with different gas chromatographic columns, operate simultaneously on the ISS to accurately measure target compounds important to the assessment of the onboard air quality. The AQMs had to be validated, which meant they had to demonstrate their ability to accurately analyze target compounds in the ISS atmosphere, before they could be proclaimed ready for operational use. As the name implies, operational hardware can and will be used to make real-time decisions; therefore validation is necessary to have confidence in generated data. The AQMs were deployed in late March 2013 and validation was begun in May 2013. During the in-flight validation phase, a minimum of 6 mini-grab sample containers (mini-GSCs) were acquired nearly simultaneously with the AQM sample analysis. Six months was needed to acquire the necessary mini-GSC-AQM pairings to complete validation. The validation criteria, established prior to launching the AQMs, were applied to each compound and not to the AQM units. Furthermore, the validation included the checkout of the AQM battery operations in a remote location. The AQMs operate autonomously using a scripted analysis scheme and frequency. A remote desktop feature of the AQMs permits the scripts to be interrupted, without use of crew time, so that the AQM runs could be coordinated with the mini-GSC sample acquisition. This paper will briefly describe the AQM technology, describe the validation requirements, and present the results from the AQM validation.en_US
dc.format.mimetypeapplication/pdf
dc.identifier.isbn978-0-692-38220-2
dc.identifier.otherICES-2014-218
dc.identifier.urihttp://hdl.handle.net/2346/59723
dc.language.isoengen_US
dc.publisher44th International Conference on Environmental Systemsen_US
dc.titleOperational Validation of the Air Quality Monitor on the International Space Stationen_US
dc.typePresentationen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
ICES-2014-218.pdf
Size:
317.36 KB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.57 KB
Format:
Item-specific license agreed upon to submission
Description: